Application of Glow Curve Analysis Methods to Improve TLD-100 Dose Reassessment Performance

Abstract
This paper presents a method of improving the TLD-100 dose reassessment performance. This method consists of applying numerical analysis techniques for evaluating the TLD-100 phototransferred thermoluminescence (PTTL) glow curve. From this analysis, a simple procedure for estimating the ultraviolet background components usually present in phototransferred thermoluminescence (TL) signals has been established. This procedure has been implemented in a computer program which performs the automatic evaluation of the glow curves and extracts the dose information contained in the PTTL curves. The use of this computer-aided evaluational method has enabled the extension of the working range of estimated absorbed dose down to 0.2-0.5 mGy with very adequate operational quality for doses even below the conventionally admitted lower reestimation limit (approximately 2 mGy). Because TL readout is a destructive process, the ability to reestimate doses can be important in any kind of dosimetric activity, such as operational dosimetry programs. The other commonly used dosimeter, film, uses a nondestructive readout and, therefore, presents some advantages over TLD when dose reassessment is necessary. With the reported improvements in the TLD-100 dose reassessment performance, the full range of absorbed doses covered by film dosimetry can now be reliably reassessed using TLD-100 dosimeters.

This publication has 0 references indexed in Scilit: