A portable versatile X- and gamma -ray dosemeter with digital display for medical use

Abstract
A multipurpose dosemeter based on analogue-to-digital (A-D) conversion has been developed using solid state devices, micropower integrated circuits and displays. The A-D converter combines the features of a MOSFET electrometer, a voltage-to-frequency converter, the automatic Townsend balance, LED(LCD) display for exposure, analogue output and has a voltage-measuring accuracy of +or-0.1% FS+or-1 digit. The dosemeter, which is designed to measure a wide range of X- and gamma -ray exposure rates (1 mRh-1 to 1000 R min-1) and exposures (1 mR to 1000 R) with a variety of air-equivalent ionisation chambers, is compact, rugged and battery or AC mains powered. It is designed to provide trouble-free performance, easy operation and maintenance and to be independent of climatic conditions. Combining the features of a survey meter, dosemeter and a radioisotope calibrator, it can be fabricated using the facilities locally available in hospitals or universities.

This publication has 0 references indexed in Scilit: