Correction for Non-Linearity of Proportional Counter Systems in Electron Probe X-Ray Microanalysis
- 1 January 1965
- journal article
- Published by Cambridge University Press (CUP) in Advances in X-ray Analysis
- Vol. 9, 208-220
- https://doi.org/10.1154/s0376030800003554
Abstract
While the theoretical basis for the correction of non-linearity of detector systems is well known, methods for the determination of dead-time effects must be adapted to electron probe microanalyzer systems. Two such methods, one employing both X-ray and current measurements and the other employing simultaneous X-ray measurements on two spectrometers, are described. The effect of pulse-height shrinkage at high counting rates on the linearity of the detector system is discussed. When the proposed corrections for the dead-time of X-ray detector systems employing proportional counters are applied to the X-ray intensity measurements obtained with the electron probe microanalyzer, count rates as high as 50,000 counts/sec can be used.Keywords