IQ™ Technology: Development of a Universal, Homogeneous Method for High-Throughput Screening of Kinase and Phosphatase Activity
Open Access
- 1 April 2003
- journal article
- Published by SAGE Publications in SLAS Technology
- Vol. 8 (2) , 36-40
- https://doi.org/10.1016/s1535-5535(04)00249-7
Abstract
IQ™ Technology is a homogeneous, universal detection platform designed specifically for high-throughput screening of kinases and phosphatases. The technology has been tested in 96- to 384- to 1536-well microplate formats and is ideally suited for automated drug discovery screening systems. IQ™ Technology is a direct, non-competitive assay format that does not require antibodies or radioactive reagents to measure phosphorylation state. Kinase or phosphatase activity is quantitated by direct measurement of the phosphorylation state of fluorophore-labeled peptide substrates. Phosphorylation is measured by the change in fluorescence intensity that occurs when a proprietary iron-containing compound binds specifically to phosphoryl groups on peptides. This change in observed fluorescence is proportional to the extent of phosphorylation of the fluorophore-labeled peptide, because the iron-containing compound quenches the fluorescent emission exclusively on the phosphorylated form of the fluorophore-labeled peptide. The IQ™ Technology provides a universal method that can be used with any peptide sequence.Keywords