Some observations of the design of rf coils for human internal use
- 1 January 1988
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 6 (1) , 49-62
- https://doi.org/10.1002/mrm.1910060106
Abstract
Some features of the design and implementation of NMR probes for insertion into peripheral orifices of the body are described. In particular, the possibilities of using parasitic coils without direct connections in conjunction with conventional whole body coils are examined. The use of windings around the NMR probe carrying gradient pulses as a means of eliminating the very high signals from tissue immediately adjacent to the coil windings is also evaluated. These approaches are, in practice, difficult to combine, and hence need to be viewed almost as alternative possibilities. Practical results of experimental investigations are presented using both phantoms and volunteers. © 1988 Academic Press, Inc.This publication has 5 references indexed in Scilit:
- Wireless implanted magnetic resonance probes for in vivo NMRJournal of Magnetic Resonance (1969), 1986
- Inductive (flux linkage) coupling to local coils in magnetic resonance imaging and spectroscopyJournal of Magnetic Resonance (1969), 1986
- Use of Closely Coupled Receiver Coils in MR ImagingJournal of Computer Assisted Tomography, 1985
- Initial Clinical Evaluation of a Whole Body Nuclear Magnetic Resonance (NMR) TomographJournal of Computer Assisted Tomography, 1982
- Mapping of metabolites in whole animals by 31P NMR using surface coilsNature, 1980