A rapid micro method for recording red cell osmotic fragility by continuous decrease of salt concentration
Open Access
- 1 July 1963
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 16 (4) , 377-382
- https://doi.org/10.1136/jcp.16.4.377
Abstract
A single volume of 0·075 ml. of a 1 in 10 dilution of whole blood in isotonic NaCl solution is introduced into a container cell, two walls of which are made of a dialysing membrane. The container cell is introduced into a test tube of distilled water, placed in an instrument which is essentially a colorimeter with a recorder, between the source of light and the photoelectric cell. Dialysis through the membrane results in a continuous decrease in the salt concentration of the medium surrounding the erythrocytes. The measurement of the degree of haemolysis is based on the increasing transparency of the erythrocyte suspension while haemolysis takes place. Recording this increasing light transmission as a function of time,i.e.,as a function of decreasing salt concentration, yields the osmotic fragility curve. The automatically recorded curve is obtained in less than 10 minutes.Keywords
This publication has 4 references indexed in Scilit:
- An Instrument for Automatically Recording the Osmotic Fragility Curve of Red Cells and/or Its DerivativeIRE Transactions on Bio-Medical Electronics, 1963
- Osmotic hemolysis by a gradual decrease in the ionic strength of the surrounding mediumJournal of Cellular and Comparative Physiology, 1961
- THE DISTRIBUTION CURVE OF ERYTHROCYTE FRAGILITYBlood, 1949
- The estimation of the fragility of red blood corpusclesThe Journal of Pathology and Bacteriology, 1938