The Energy Response of a Victoreen Condenser Ionization Chamber
- 1 January 1959
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 3 (3) , 207-211
- https://doi.org/10.1088/0031-9155/3/3/302
Abstract
The energy response of a graphite wall chamber and a 25-r typical Victoreen chamber over the range of photon energies from 40 kev to 1.2 mev were measured and compared. In the conventional X-ray range, the response of the Victoreen chamber does not differ by more than about 2% from a mean value of 1.08; but between the upper end of the X-ray region at 100 kev and the flat region at higher energies there is a drop of over 6% in the response. Theoretical response curves for chambers composed of combinations of graphite and heavier elements are compared with the Victoreen curve.Keywords
This publication has 3 references indexed in Scilit:
- MODIFICATION OF THE BASIS FOR ROENTGEN CALIBRATIONS BETWEEN 0.5 and 3 Mev.Canadian Journal of Physics, 1958
- A Comparison of X-ray StandardsThe British Journal of Radiology, 1954
- ber das Grundgesetz der physikalischen Wirkungen von R ntgenstrahlen verschiedener Wellenl nge. IIThe European Physical Journal A, 1953