An electrical caliper for continuous measurement of relative displacement
- 1 January 1962
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 17 (1) , 131-134
- https://doi.org/10.1152/jappl.1962.17.1.131
Abstract
An electrical recording caliper for the continuous measurement of relative displacement has been developed. The details of design, operation, and performance are described. The frequency response was uniform (±5%) through 25 cycle/sec. An approximately 0.5% error related to the mechanical impedance of the device was estimated to occur. The system is linear and free of hysteresis. Base-line and gain stability are acceptable. Submitted on July 21, 1961Keywords
This publication has 1 reference indexed in Scilit: