Dialysate Flow Distribution in Hollow Fiber Hemodialyzers with Different Dialysate Pathway Configurations
- 1 September 2000
- journal article
- research article
- Published by SAGE Publications in The International Journal of Artificial Organs
- Vol. 23 (9) , 601-609
- https://doi.org/10.1177/039139880002300902
Abstract
The efficiency of a hemodialyzer is largely dependent on its ability to facilitate diffusion, since this is the main mechanism by which small solutes are removed. The diffusion process can be impaired if there is a mismatch between blood and dialysate flow distribution in the dialyzer. The objective of the paper was to study the impact of different dialysate compartment designs on dialysate flow distribution and urea clearances. Eighteen hollow fiber 1.3 m2 hemodialyzers were studied, 6 each of 3 designs: Type A- standard fiber bundle (PAN 65DX Asahi Medical, Tokyo, Japan); Type B - spacing filaments external to the fibers (PAN 65SF Asahi Medical, Tokyo, Japan); Type C - fibers waved to give Moiré structure (FB130 Nissho-Nipro, Osaka, Japan). In vitro studies: 3 dialyzers of each type were studied following dye injection into the dialysate compartment. Dynamic sequential imaging of longitudinal sections of the dialyzer were undertaken, using a new generation helical CT scanner (X-Press/HS1 Toshiba Corporation, Tokyo, Japan). In vivo studies: 3 dialyzers of each type were studied, in randomized sequence, in 3 different patients under standardized dialysis conditions. Blood- and dialysate-side urea clearances were measured at 30 and 150 minutes of treatment. Macroscopic and densitometrical analysis revealed that flow distribution was most homogeneous in the dialyzer with Moiré structure (Type C) and least homogeneous in the standard dialyzer (Type A). Space yarns (Type B) gave an intermediate dialysate flow distribution. Significantly increased urea clearances (p<0.001) were seen with Types B and C, compared to the standard dialyzer. Type C (Moiré) had the highest clearances although these were not significantly greater than Type B (space yarns). In conclusion, more homogeneous dialysate flow distribution and improved small solute clearances can be achieved by use of spacing yarns or waved (Moiré structure) patterns of fiber packing in the dialyzer. These effects are achieved probably as a result of reduced dialysate channeling resulting in a lower degree of mismatch between blood and dialysate flows. The new radiological technique using the helical CT scanner allows detailed flow distribution analysis and has the potential for testing future modifications to dialyzer design.Keywords
This publication has 18 references indexed in Scilit:
- A new scintigraphic method to characterize ultrafiltration in hollow fiber dialyzersKidney International, 1992
- Ultrafiltration and Pressure Profiles in Continuous Arteriovenous Hemofiltration Studied by Computerized Scintigraphic ImagingThe International Journal of Artificial Organs, 1991
- Importance of Hollow-Fiber Geometry in Continuous Arteriovenous HemofiltrationPublished by S. Karger AG ,1990
- Impact of High Blood Flows on Vascular Stability in HaemodialysisNephrology Dialysis Transplantation, 1990
- Hollow‐Fiber Dialyzers on the MarketArtificial Organs, 1988
- Theoretical assessment of the effect of flow maldistributions on the mass transfer efficiency of artificial organsMedical & Biological Engineering & Computing, 1985
- Comparison of Hollow Fiber DialyzersArtificial Organs, 1981
- HEMODIALYZER EVALUATION BY EXAMINATION OF SOLUTE MOLECULAR SPECTRAAsaio Journal, 1972
- A Totally Disposable, Presterilized DialyzerArchives of internal medicine (1960), 1971
- Dialysieren des Strömenden Blutes am LebendenJournal of Molecular Medicine, 1923