Performance of an immunochromatographic test for the rapid diagnosis of malaria
- 1 April 2004
- journal article
- research article
- Published by Springer Nature in Zeitschrift Fur Parasitenkunde-Parasitology Research
- Vol. 92 (6) , 518-519
- https://doi.org/10.1007/s00436-004-1069-9
Abstract
The immunochromatographic test (ICT) for the rapid diagnosis of malaria has been marketed for several years. In a study in which three Centres of Tropical Medicine participated and data were pooled, performance of the test varied considerably when comparing the results between each centre. The sensitivity of ICT in 2,343 patients tested in our services was 100% and the specificity 99.74%. Moreover, two patients with a positive ICT would initially have been missed by expert microscopy, with Plasmodium falciparum malaria being confirmed microscopically some hours later. The principal reasons for the better performance of the test in our series appear to be blood collection in EDTA vials and considerable experience with handling and interpreting the ICT test.Keywords
This publication has 2 references indexed in Scilit:
- Comparison of three antigen detection tests for diagnosis and follow-up of falciparum malaria in travellers returning to Berlin, GermanyZeitschrift Fur Parasitenkunde-Parasitology Research, 2003
- Evaluation of a Dipstick Test for the Rapid Diagnosis of Imported Malaria Among Patients Presenting Within the Network TropNetEuropScandinavian Journal of Infectious Diseases, 2001