Physical comparison of gamma camera and two-headed scanner: figures of merit and times required for threshold detection of spherical sources
- 1 August 1968
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 41 (488) , 586-596
- https://doi.org/10.1259/0007-1285-41-488-586
Abstract
Physical comparisons are described between a scintillation camera and a 2-headed scanner for their efficiency in detecting radioactive spheres in tanks of water containing radioactivity at different concentrations. For these comparisons, the previously defined figure of merit has to be modified to use "counts recorded per integratng time, i.e. time spent over the abnormal region" instead of "per minute". A new quantity, "merit-time-product", is defined and used for this purpose. It is then possible to compare scanners moving at different speeds with each other or with cameras, on the same basis. Results are presented showing the total times required for each device to scan an area, so that a given level of significance is achieved in the threshold detection of active spheres in the presence of background radioactivity. The scintillation camera requires shorter times. The differences are small for 131L but exceed factors of 20 in overall time for 99mTc. These factors decrease with depth in the scintillation camera, from 20 to 30 at 7 cm to about 10 at 10 cm depth. Possibly the most striking difference between the camera and the 2-head scanner is the high efficiency of the camera for relatively shallow tumors and the rapid decrease of efficiency with depth.This publication has 10 references indexed in Scilit:
- The effect of collimator resolution on the detection of lesions in brain scanningThe British Journal of Radiology, 1968
- The response of double-head and single-head collimators for scanningThe International Journal of Applied Radiation and Isotopes, 1968
- The longitudinal and lateral response of multichannel focusing collimatorsThe International Journal of Applied Radiation and Isotopes, 1968
- An Investigation into the Usefulness of "Figure of Merit" as a Criterion of a Collimating SystemPhysics in Medicine & Biology, 1967
- The Quantitative Analysis of Cerebral ScintiscansPhysics in Medicine & Biology, 1966
- Radioisotope Scanning DisplayPhysics in Medicine & Biology, 1965
- Comparison of Coincidence Counting and Focusing Collimators with Various Isotopes in Brain Tumour DetectionThe British Journal of Radiology, 1964
- Criteria for evaluating collimators used in in vivo distribution studies with radioisotopesThe International Journal of Applied Radiation and Isotopes, 1961
- IV. A Quantitative Automatic Body Scanner for the Localisation of Radioisotopesin vivoThe British Journal of Radiology, 1959
- The Sensitivity Performance of the Human Eye on an Absolute Scale*Journal of the Optical Society of America, 1948