A scheme for accidental coincidence correction in time-of-flight positron tomography: theory and implementation
- 1 February 1988
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 35 (1) , 753-756
- https://doi.org/10.1109/23.12826
Abstract
A computationally inexpensive method is described for correcting time-of-flight positron-emission-tomography (PET) images for accidental coincidences. Initial runs of the algorithm show that it rapidly (2-6 iterations) yields accurate estimates of single rates from simulated data corrupted by Poisson noise. Two illustrative experiments with the method are described.Keywords
This publication has 3 references indexed in Scilit:
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1986
- Image-Reconstruction of Data from Super PETT I: A First-Generation Time-of-Flight Positron-Emission TomographIEEE Transactions on Nuclear Science, 1986