Automated detection of malaria by means of the haematology analyser CoulterR GEN.STM
- 1 December 2004
- journal article
- Published by Wiley in Clinical and Laboratory Haematology
- Vol. 26 (6) , 367-372
- https://doi.org/10.1111/j.1365-2257.2004.00648.x
Abstract
The haematology analysers Coulter(R) GEN.S(TM) and LH(TM) give a set of data--'positional parameters'--which define each WBC population by mean of index values, the mean and the standard deviation (SD) of volume, conductivity and scatter, used to identify the WBC populations. These parameters were analysed in patients investigated for suspicion of malaria, in order to show a difference between malaria negative and malaria positive patients and to use it for malaria detection. The six parameters exhibiting a significant difference between the two groups were submitted to a ROC analysis, which showed both sensitivity >90% and specificity >60% for two parameters, lymphocyte and monocyte SDs of volumes. A discriminator combining the two parameters showed a sensitivity of 96.9% and a specificity of 82.5%. The cut off of the discriminative value was calculated. Because of the good stability and reproducibility of the parameters selected, the test can be used in order to detect patients having a high probability to be malaria positive and to pay particular attention to these blood smears. For more extensive diagnostic use, a standard control procedure of the positional parameters should be introduced.Keywords
This publication has 7 references indexed in Scilit:
- Evaluation of the coulter LH 750 haematology analyzer compared with flow cytometry as the reference method for WBC, platelet and nucleated RBC countInternational Journal of Laboratory Hematology, 2004
- Sensitivity of hemozoin detection by automated flow cytometry in non‐ and semi‐immune malaria patientsCytometry Part B: Clinical Cytometry, 2003
- Automated detection of malaria-associated intraleucocytic haemozoin by Cell-Dyn CD4000 depolarization analysisClinical and Laboratory Haematology, 2003
- Relative frequency of malaria pigment-carrying monocytes of nonimmune and semi-immune patients from flow cytometric depolarized side scatterCytometry, 2001
- Automated malaria detection by depolarization of laser lightBritish Journal of Haematology, 1999
- Intraleucocytic malaria pigment and clinical severity of malaria in childrenTransactions of the Royal Society of Tropical Medicine and Hygiene, 1998
- Malarial pigment (haemozoin): a very active ‘inert’ substancePathogens and Global Health, 1997