An experimental investigation on intra-fractional organ motion effects in lung IMRT treatments
Top Cited Papers
- 3 June 2003
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 48 (12) , 1773-1784
- https://doi.org/10.1088/0031-9155/48/12/307
Abstract
Respiration-induced tumour motion can potentially compromise the use of intensity-modulated radiotherapy (IMRT) as a dose escalation tool for lung tumour treatment. We have experimentally investigated the intra-fractional organ motion effects in lung IMRT treatments delivered by multi-leaf collimator (MLC). An in-house made motor-driven platform, which moves sinusoidally with an amplitude of 1 cm and a period of 4 s, was used to mimic tumour motion. Tumour motion was simulated along cranial-caudal direction while MLC leaves moved across the patient from left to right, as in most clinical cases. The dose to a point near the centre of the tumour mass was measured according to geometric and dosimetric parameters from two five-field lung IMRT plans. For each field, measurement was done for two dose rates (300 and 500 MU min−1), three MLC delivery modes (sliding window, step-and-shoot with 10 and 20 intensity levels) and eight equally spaced starting phases of tumour motion. The dose to the measurement point delivered from all five fields was derived for both a single fraction and 30 fractions by randomly sampling from measured dose values of each field at different initial phases. It was found that the mean dose to a moving tumour differs slightly (<2–3%) from that to a static tumour. The variation in breathing phase at the start of dose delivery results in a maximum variation around the mean dose of greater than 30% for one field. The full width at half maximum for the probability distribution of the point dose is up to 8% for all five fields in a single fraction, but less than 1–2% after 30 fractions. In general, lower dose rate can reduce the motion-caused dose variation and therefore might be preferable for lung IMRT when no motion mitigation techniques are used. From the two IMRT cases we studied where tumour motion is perpendicular to MLC leaf motion, the dose variation was found to be insensitive to the MLC delivery mode.Keywords
This publication has 12 references indexed in Scilit:
- Synchronized moving aperture radiation therapy (SMART): average tumour trajectory for lung patientsPhysics in Medicine & Biology, 2003
- Intraindividual Comparison of Conventional Three-Dimensional Radiotherapy and Intensity Modulated Radiotherapy in the Therapy of Locally Advanced Non-Small Cell Lung CancerStrahlentherapie und Onkologie, 2002
- Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapyInternational Journal of Radiation Oncology*Biology*Physics, 2002
- Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulationPhysics in Medicine & Biology, 2002
- Issues in respiratory motion compensation during external-beam radiotherapyInternational Journal of Radiation Oncology*Biology*Physics, 2002
- Delivery of intensity‐modulated radiation therapy with a conventional multileaf collimator: Comparison of dynamic and segmental methodsMedical Physics, 2001
- Influence of respiration‐induced organ motion on dose distributions in treatments using enhanced dynamic wedgesMedical Physics, 2001
- Intensity-modulated radiotherapy: current status and issues of interestInternational Journal of Radiation Oncology*Biology*Physics, 2001
- A new approach to dose escalation in non–small-cell lung cancerInternational Journal of Radiation Oncology*Biology*Physics, 2001
- Improved survival of patients with unresectable non-small-cell bronchogenic carcinoma by an innovated high-doseEn-Bloc radiotherapeutic approachCancer, 1981