Clinical evaluation and comparison of the Infrasonde, Arteriosonde, and mercury sphygmomanometer in measurement of blood pressure in children.
Open Access
- 28 February 1979
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood
- Vol. 54 (3) , 184-189
- https://doi.org/10.1136/adc.54.3.184
Abstract
Both systolic and diastolic pressures can be measured in children with the mercury sphygmomanometer, the Arteriosonde and the Infrasonde. Estimates made blindly with these instruments were compared with directly measured intra-arterial blood pressures in 50 children aged between 4 days and 14 years. Systolic and diastolic estimates with the three techniques showed highly significant correlations with simultaneous intra-arterial measurements (P less than 0.001). The Infrasonde diastolic estimates were least satisfactory and the slope of the regression line against the intra-arterial pressure differed significantly from unity (y=0.54x+29.53). In 11 small children a satisfactory diastolic estimate could not be obtained with the mercury sphygmomanometer. While the mercury sphygmomanometer should remain the standard hospital equipment, an Arteriosonde would be valuable if it is difficult to hear Korotkoff's sounds on auscultation and if a diastolic pressure is required. For research investigations into childhood blood pressure an Arteriosonde or mercury sphygmomanometer, coupled with a device to exclude observer bias, is probably most suitable. Although the Infrasonde is not sufficiently accurate for research purposes, it is acceptable for routine ward use.This publication has 10 references indexed in Scilit:
- Blood pressure and its significance in childhoodPublished by Oxford University Press (OUP) ,1978
- Indirect Methods of Blood Pressure MeasurementPediatric Clinics of North America, 1978
- THE INFRASOUND BLOOD-PRESSURE RECORDER A CLINICAL EVALUATIONThe Lancet, 1976
- Assessment of Doppler ultrasound to measure systolic and diastolic blood pressures in infants and young children.Archives of Disease in Childhood, 1975
- Assessment of techniques for measurement of blood pressure in infants and children.Archives of Disease in Childhood, 1973
- Systemic Hypertension in Children and AdolescentsPediatric Clinics of North America, 1971
- MEASUREMENT OF BLOOD PRESSURE IN INFANTS AND CHILDREN BY THE DOPPLER ULTRASONIC TECHNIQUEPediatrics, 1971
- Hypertension in apparently normal childrenThe Journal of Pediatrics, 1971
- Measurement of Blood-pressure: Comparison of Intra-arterial and Cuff ValuesBMJ, 1964