Operative assessment of femorodistal bypass grafts using a new Doppler flowmeter
- 1 September 1989
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Surgery
- Vol. 76 (9) , 925-928
- https://doi.org/10.1002/bjs.1800760917
Abstract
Operative measurements using a newly developed Doppler flowmeter were made on 47 in situ femorodistal bypass grafts performed for critical ischaemia. Graft blood flow, peripheral resistance and impedance were measured before and after the injection of 15 mg papaverine down the graft. The proportion of retrograde blood flow at the distal anastomosis was also measured. The success of a graft at 1 month was defined by patency on duplex ultrasound scanning, a rise in the Doppler ankle pressures and an improvement in the clinical state of the limb. By these criteria the 1-month primary success rate was 35 out of 47 (74 per cent). Graft blood flow was of little value in predicting graft outcome but, combined with segmental pressures, accurately located six of eight technical problems which were successfully corrected. After papaverine, the peripheral resistance of all the successful grafts fell below one peripheral resistance unit (PRU). No single measurement was able to completely distinguish between success and failure, although discriminant analysis of all measurements was able to predict success with a sensitivity of 100 per cent and specificity of 97 per cent. A simpler combination of a peripheral resistance after papaverine <1 PRU and a retrograde blood flow <33 per cent predicted success, with a sensitivity of 97 per cent and specificity 83 per cent. These criteria have been validated on a further 26 grafts, achieving a sensitivity of 88 per cent and a specificity of 75 per cent. Simple operative measurements, using a newly developed Doppler flowmeter, enable the detection of technical problems and the accurate prediction of the early outcome of femorodistal grafting.Keywords
This publication has 14 references indexed in Scilit:
- The non-reversed vein femoro-distal bypass graft: A modification of the standard in situ techniqueEuropean Journal of Vascular Surgery, 1989
- A Doppler flowmeter for use in theatreUltrasound in Medicine & Biology, 1986
- Suggested standards for reports dealing with lower extremity ischemiaJournal of Vascular Surgery, 1986
- Peripheral resistance measurement in the assessment of severe peripheral vascular diseaseBritish Journal of Surgery, 1985
- Measurement of blood flow by ultrasound: Accuracy and sources of errorUltrasound in Medicine & Biology, 1985
- The value of perioperative Doppler ankle systolic pressure measurements in reconstructive arterial surgeryBritish Journal of Surgery, 1985
- Blood flow and blood velocity measurementin vivo by electromagnetic inductionMedical & Biological Engineering & Computing, 1984
- Stenosis of the common bile duct in chronic pancreatitisBritish Journal of Surgery, 1982
- The hydraulic impedance of the lower limb: Its relevance to the success of bypass operations for occlusion of the superficial femoral arteryBritish Journal of Surgery, 1976
- PROGNOSTIC VALUE OF INTRAOPERATIVE BLOOD-FLOW MEASUREMENTS IN FEMOROPOPLITEAL BYPASS VEIN-GRAFTSThe Lancet, 1968