Comparison of an Industry-Derived LCx Chlamydia pneumoniae PCR Research Kit to In-House Assays Performed in Five Laboratories
Open Access
- 1 July 2002
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 40 (7) , 2357-2362
- https://doi.org/10.1128/jcm.40.7.2357-2362.2002
Abstract
In a multicenter comparison of PCR assays utilizing 120 quantitated samples of 16 Chlamydia pneumoniae isolates, an LCx research-use-only (RUO) PCR developed by Abbott Laboratories demonstrated 100% sensitivity on 48 samples with >1 copy of DNA per μl of specimen. The sensitivities of five in-house PCR assays ranged from 54 to 94% for the same samples. All six assays showed decreased sensitivities as the DNA copy numbers of the samples decreased. Overall, sensitivities ranged from 68% for the LCx PCR assay to 29% for one of the in-house tests. The LCx RUO PCR and three of the five in-house PCR tests reported no false positives with the 24 negative samples. Increasing the number of replicates tested increased the sensitivities of all of the assays, including the LCx PCR. The LCx RUO assay showed high reproducibility for a single technologist and between technologists, with a kappa agreement of 0.77. The within-center agreements of the five in-house PCR tests varied from 0.19 to 0.74 on two challenges of 60 specimens 1 month apart. The LCx C. pneumoniae RUO PCR shows excellent potential for use in clinical studies, which could enable standardization of results in the field.Keywords
This publication has 18 references indexed in Scilit:
- StandardizingChlamydia pneumoniaeAssays: Recommendations from the Centers for Disease Control and Prevention (USA) and the Laboratory Centre for Disease Control (Canada)Clinical Infectious Diseases, 2001
- Replicate PCR Testing and Probit Analysis for Detection and Quantitation of Chlamydia pneumoniae in Clinical SpecimensJournal of Clinical Microbiology, 2001
- Circulating Nucleic Acids of Chlamydia pneumoniae and Cytomegalovirus in Patients Undergoing Coronary AngiographyJournal of Clinical Microbiology, 2001
- Comparison of two serological methods and a polymerase chain reaction-enzyme immunoassay for the diagnosis of acute respiratory infections with Chlamydia pneumoniae in adultsJournal of Medical Microbiology, 1998
- High Incidence of Chlamydia pneumoniae in Sclerotic Heart Valves of Patients Undergoing Aortic Valve ReplacementScandinavian Journal of Infectious Diseases, 1997
- The nPCR detection of Chlamydia pneumoniae and Chlamydia trachomatis in children hospitalized for bronchiolitisJournal of Infection, 1996
- Development of a simplified polymerase chain reactionenzyme immunoassay for the detection of Chlamydia pneumoniaeJournal of Applied Bacteriology, 1996
- Detection of Chlamydia pneumoniae TWAR in Human Coronary Atherectomy TissuesThe Journal of Infectious Diseases, 1995
- Detection of Chlamydia pneumoniae using a general Chlamydia polymerase chain reaction with species differentiation after hybridisationJournal of Microbiological Methods, 1993
- Detection of Chlamydia pneumoniae and Chlamydia psittaci in sputum samples by PCR.Journal of Clinical Pathology, 1993