Column Performance and Stability for High-Speed Vacuum-Outlet GC of Volatile Organic Compounds Using Atmospheric Pressure Air as Carrier Gas
- 14 October 1999
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 71 (22) , 5199-5205
- https://doi.org/10.1021/ac990573y
Abstract
The development of lightweight, portable GC instrumentation is handicapped by the need for compressed carrier gas to drive the separation. The use of air as carrier gas eliminates the need for compressed gas tanks. If a vacuum pump is used to pull carrier gas and injected samples through the column, atmospheric pressure air can be used as carrier gas. Vacuum outlet operation also improves performance for high-speed separations by reducing detector dead time and by shifting optimal carrier gas velocity to higher values. Under vacuum outlet conditions using atmospheric pressure air as carrier gas, a 6-m-long, 0.25-mm-i.d. capillary column can generate ∼12 500 theoretical plates, and a 12-m-long column can generate ∼44 000 plates but with a 3−4-fold increase in separation time. The principal issues in column selection for high-speed GC with air as a carrier gas are efficiency and stability. Several bonded and nonbonded stationary phases were evaluated for use with air as carrier gas in the analysis of volatile organic compounds of interest in air-monitoring applications. These include dimethylpolysiloxane, 50% phenyl−50% methyl polysiloxane, 50% cycanopropylphenyl−50% methyl polysiloxane, trifluropropyl polysiloxane, poly(ethylene glycol), and dicyanoallyl polysiloxane (nonbonded). The dimethyl polysiloxane and the trifluropropyl polysiloxane columns showed good efficiency and no significant deterioration after 5 days of continuous operation with air as carrier gas. The 50% phenyl−50% methyl polysiloxane and the 50% cycanopropylphenyl−50% methyl polysiloxane columns showed poorer efficiency, and the poly(ethylene glycol) and dicyanoallyl polysiloxane columns showed excessive deterioration in air.Keywords
This publication has 19 references indexed in Scilit:
- Peer Reviewed: High-speed gas chromatographyAnalytical Chemistry, 1998
- Novel Column Heater for Fast Capillary Gas ChromatographyJournal of Chromatographic Science, 1996
- High-speed GC analysis of VOCs: sample collection and inlet systems. 1.Environmental Science & Technology, 1994
- High‐speed gas chromatography using an on‐column thermal desorption modulatorJournal of Microcolumn Separations, 1989
- Sample introduction in high speed capillary gas chromatography; input band width and detection limitsJournal of High Resolution Chromatography, 1987
- Optimum performance of capillary GC columns as a function of tube diameter and film thickness under various operating conditions. Computer program for calculation of H‐ū curves and minimum analysis timesJournal of High Resolution Chromatography, 1985
- Silicon Micromechanical DevicesScientific American, 1983
- Increased speed of analysis in directly coupled gas chromatography-mass spectrometry systemsJournal of Chromatography A, 1982
- Increased speed of analysis in directly coupled gas chromatography-mass spectrometry systems : Capillary columns at sub-atmospheric outlet pressuresJournal of Chromatography A, 1981
- Influence of instrumental contributions on the apparent column efficiency in high speed gas chromatographyAnalytical Chemistry, 1978