Method of measuring loaded Q-factor of single-ended cavity resonators using reflection bridge
- 1 March 1978
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 49 (3) , 378-379
- https://doi.org/10.1063/1.1135411
Abstract
A bridge method of determining the loaded Q-factor and coupling coefficient of single-ended cavity resonators is described. The measuring system consists of a hybrid-T or ’’magic T’’ bridge with one of the collinear arms terminated by the cavity and the other by a precision attenuator followed by a variable precision short. The experimental procedure essentially measures the reflection coefficient of a cavity resonator as a function of frequency by balancing the bridge, using an oscillator with variable frequency amplitude modulation as the source.Keywords
This publication has 1 reference indexed in Scilit:
- A bridge method for simultaneous measurements of coupling coefficient and loaded Q of a single-ended cavityReview of Scientific Instruments, 1973