Separation of Polybutylene Glycols on C18and C4Stationary Phases

Abstract
Polybutylene glycol (PBG) samples largely differing in molecular weight Mr are separated by gradient reversed-phase high performance liquid chromatgraphy on C18 and C4 stationary phases by use of either acetonitrile or methanol as organic modifiers and signal monitoring by means of evaporative light scattering detection. Neither acetonitrile nor methanol are sufficient for the quantitative elution of all samples from the extremely hydrophobic C18 matrix. However, the elution potency of methanol is markedly better compared with acetonitrile presumably attributable to a solubility increase of sample molecules via hydrogen bonding between its hydroxy group and the ether oxygens of the solutes. Marked lower retention of high Mr oligomers is observed on the more polar C4 matrix and the elution power of acetonitrile is now sufficient for the quantitative release of all investigated polyether samples. A substantial increase of signals eluting at higher retention times and thus representing oligomers with higher Mr becomes evident in the range PBG 650 < PBG 1000 < PBG 2000 < PBG 3000. Although the PBGs exhibit a broad “within-sample” oligomeric Mr distribution and thus substantial peak overlapping by superposition of the individual chromatograms a clear assignment to individual samples can be done on the basis of the different chromatographic patterns.