The wear of ultra-high molecular weight polyethylene sliding on metallic and ceramic counterfaces representative of current femoral surfaces in joint replacement
- 1 January 1997
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 211 (1) , 17-24
- https://doi.org/10.1243/0954411971534647
Abstract
A number of studies have investigated the influence of surface roughness on the wear of ultra-high molecular weight polyethylene (UHMWPE) in total joint replacement. The results of these studies have shown that the wear factor is proportional to the counterface roughness raised to a power greater than one. In this laboratory study, the effect of surface finish of several biomaterials on the wear of UHMWPE was studied. The study was conducted using reciprocating pin-on-plate wear tests with bovine serum as a lubricant. The biomaterials investigated as the counterface material included stainless steel, cast cobalt chrome (CoCr), CoCr (ASTM F799), alumina ceramic and zirconia ceramic. The counterface topographies of the wear plates were produced using techniques representative of current manufacturing methods. The surface roughness of the wear plates was varied in the range Ra, = 0.005-0.04 μm; this was representative of femoral heads and femoral knee components currently used clinically. Metals and ceramics with a similar surface roughness produced a similar wear rate of UHMWPE. For the limited range of smooth counterfaces used in this study only a moderate correlation was found between the surface roughness and the wear factors. For a change in counterface roughness Ra of 0.005 to 0.04 μm, the wear factor increased from 7.4 ± 1.6 to 16.5 ± 2.4 × 10−9mm3/N m (mean ± standard error). This variation in counterface roughness had much less effect in wear than previously reported for rougher counterfaces. For an extended range of counterface roughness, a stronger correlation was found using an exponential function for the regression fit. The exponential function shows the benefits of decreased wear with decreased surface roughness. Although the wear rate decreased less rapidly with decreased counterface roughness for Ra values below 0.05 μa, there were significant advantages to be gained from improved femoral head roughness to below 0.01 μm RaKeywords
This publication has 27 references indexed in Scilit:
- The effect of transfer film and surface roughness on the wear of lubricated ultra-high molecular weight polyethyleneClinical Materials, 1993
- Wear of the polyethylene acetabular cup: Metallic and ceramic heads compared in a hip simulatorActa Orthopaedica, 1993
- Low wear rate of UHMWPE against zirconia ceramic (Y‐PSZ) in comparison to alumina ceramic and SUS 316L alloyJournal of Biomedical Materials Research, 1991
- Ion implantation effects on friction and wear of joint prosthesis materialsBiomaterials, 1991
- EditorialEngineering in Medicine, 1988
- EditorialEngineering in Medicine, 1988
- The effect of the surface finish of alumina and stainless steel on the wear rate of UHMW polyethyleneBiomaterials, 1986
- Influence of Counterface Topography on the Wear of Ultra High Molecular Weight Polyethylene Under Wet or Dry ConditionsPublished by American Chemical Society (ACS) ,1985
- Friction and wear properties of polymer, metal, and ceramic prosthetic joint materials evaluated on a multichannel screening deviceJournal of Biomedical Materials Research, 1981
- Wear characteristics of UHMW polyethylene: A method for accurately measuring extremely low wear ratesJournal of Biomedical Materials Research, 1978