Indocyanine green densitometry in flowing blood compensated for background dye
- 1 November 1963
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 18 (6) , 1294-1304
- https://doi.org/10.1152/jappl.1963.18.6.1294
Abstract
Blood is nonhomogeneous; hence, the relationship between light transmission and increasing concentration of dye in whole blood is never the perfect exponential curve predicted by Beer's law. Instead, as the concentration of indocyanine green is increased to high levels (40 mg/liter) the light transmission decreases exponentially toward an asymptote at 6–8% transmission for nearly monochromatic densitometers (half-band width: 13–20 mμ), but at 30–40% for densitometers using light of wide-band width. Consequently, following recording of a dilution curve, circulating background dye reduces the change in transmission per unit increase in dye concentration in subsequent curves. This decrease in sensitivity cannot be compensated for by a simple increase in the sensitivity of the densitometer or in the intensity of its light source. Compensation can be attained, however, if increasing densitometer sensitivity is associated with the automatic scale expansion provided when a suppressed zero point is used. At correct zero suppression, the deflection for zero output of the densitometer coincides with the asymptotic transmission value mentioned above. indicator-dilution in circulation; blood flow measurement; blood optical density; dye dilution technic; cardiac output measurement Submitted on April 15, 1963Keywords
This publication has 11 references indexed in Scilit:
- Areas of dye-dilution curves sampled simultaneously from central and peripheral sitesJournal of Applied Physiology, 1962
- INDOCYANINE GREEN - PHYSICAL AND PHYSIOLOGIC PROPERTIES1960
- HEPATIC REMOVAL OF INDOCYANINE GREEN1960
- BLOOD LEVEL OF INDOCYANINE GREEN IN THE DOG DURING MULTIPLE DYE CURVES AND ITS EFFECT ON INSTRUMENTAL CALIBRATION1960
- FACTORS DETERMINING THE DISPERSION OF DYE IN INDICATOR DILUTION CURVES IN THE NORMAL MAMMALIAN CIRCULATION1958
- Dynamic Response Characteristics of Systems for Continuous Recording of Concentration Changes in a Flowing Liquid (for Example, Indicator-Dilution Curves)Journal of Applied Physiology, 1957
- Estimations of Cardiac Output and Central Blood Volume by Dye DilutionPhysiological Reviews, 1956
- Measurement of Cardiac Output by Dye Dilution Technique: Use of an “Integrated” Sample Collection in Calibration of the Photometric InstrumentJournal of Applied Physiology, 1954
- Estimation of cardiac output and Evans blue space in man, using an oximeter.1951
- Special Instrumentation Problems Encountered in Physiological Research Concerning the Heart and Circulation in ManScience, 1950