Percentage depth doses for high energy X-rays
- 1 March 1976
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 21 (2) , 226-235
- https://doi.org/10.1088/0031-9155/21/2/004
Abstract
Percentage depth doses for 34 MV betatron X-rays were measured ionometrically for a number of square fields of side 5-20 cm. Significant differences in the depth doses for different field sizes were found in the build-up region and in the depth of maximum dose. A systematic change with depth was also indicated in the exponential region of these curves. These differences are discussed in terms of the changes in the contributions of primary radiation and of scattered radiation from both the collimator and the phantom.Keywords
This publication has 5 references indexed in Scilit:
- Physical aspects of supervoltage x-ray therapyMedical Physics, 1974
- Variation in the position of the central axis maximum build-up point with field size for high-energyThe British Journal of Radiology, 1970
- The Extension of the Concept of Tissue-Air Ratios (TAR) to High-Energy X-Ray BeamsRadiology, 1970
- Displacement Effect of Thimble Chambers Exposed to a Photon or Electron Beam from a BetatronActa Radiologica, 1967
- Etude comparée d'une série de chambres d'ionisation dans des faisceaux d'électrons de 20 et 10 MeVRadiation and Environmental Biophysics, 1966