Modelling the Femoral Load Transfer in Total Hip Joint Replacement—A Preliminary Study
- 1 April 1983
- journal article
- research article
- Published by SAGE Publications in Engineering in Medicine
- Vol. 12 (2) , 61-64
- https://doi.org/10.1243/emed_jour_1983_012_018_02
Abstract
This paper describes a simple modelling technique used to study the load transfer in the proximal femur in total hip replacement. Model femora cast in a rigid polyurethane foam and prostheses made from various plastics were used to study the effects of individual design parameters on stresses in the femur and prosthesis. The technique was applied in this study to examine the effects on the load transfer of two geometrically identical prostheses simulating two different materials, viz. titanium and chrome cobalt. The model simulating titanium produced more physiological femoral stresses and lower stem stresses. The technique proved to be a simple yet reliable method of studying individual parameters related to the design of femoral prostheses.Keywords
This publication has 4 references indexed in Scilit:
- An analysis of femoral component stem design in total hip arthroplasty.Journal of Bone and Joint Surgery, 1980
- Proximal strain distribution in the loaded femur. An in vitro comparison of the distributions in the intact femur and after insertion of different hip-replacement femoral components.Journal of Bone and Joint Surgery, 1978