Abstract
A method is described that utilizes direct current plasma atomic emission spectrometry as an element-selective method of detection for ion chromatographic determination of chromium(III) and chromium(VI) species. The eluting chromium-containing species are detected on the basis of the atomic emission of chromium without any species conversion. Both anion and cation separator columns give similar results when used with varying sample matrices. By employing an on-column preconcentration procedure, the detectable concentrations of the chromium species are reduced to less than 1.0 ppb. This method is applied to the determination of chromium species in human serum, natural water, and industrial process stream samples

This publication has 0 references indexed in Scilit: