Multi-component apparent diffusion coefficients in human brain

Abstract
The signal decay with increasing b‐factor at fixed echo time from brain tissue in vivo has been measured using a line scan Stejskal–Tanner spin echo diffusion approach in eight healthy adult volunteers. The use of a 175 ms echo time and maximum gradient strengths of 10 mT/m allowed 64 b‐factors to be sampled, ranging from 5 to 6000 s/mm2, a maximum some three times larger than that typically used for diffusion imaging. The signal decay with b‐factor over this extended range showed a decidedly non‐exponential behavior well‐suited to biexponential modeling. Statistical analyses of the fitted biexponential parameters from over 125 brain voxels (15 × 15 × 1 mm3 volume) per volunteer yielded a mean volume fraction of 0.74 which decayed with a typical apparent diffusion coefficient around 1.4 µm2/ms. The remaining fraction had an apparent diffusion coefficient of approximately 0.25 µm2/ms. Simple models which might explain the non‐exponential behavior, such as intra‐ and extracellular water compartmentation with slow exchange, appear inadequate for a complete description. For typical diffusion imaging with b‐factors below 2000 s/mm2, the standard model of monoexponential signal decay with b‐factor, apparent diffusion coefficient values around 0.7 µm2/ms, and a sensitivity to diffusion gradient direction may appear appropriate. Over a more extended but readily accessible b‐factor range, however, the complexity of brain signal decay with b‐factor increases, offering a greater parametrization of the water diffusion process for tissue characterization. Copyright © 1999 John Wiley & Sons, Ltd.