Implantation of an Automatic Defibrillator Using a New Nonthoracotomy Approach
- 1 December 1994
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 17 (12) , 2247-2254
- https://doi.org/10.1111/j.1540-8159.1994.tb02372.x
Abstract
Most current nonthoracotomy systems for defibrillator implantation use monophasic devices. To determine the safety and efficacy of a new nonthoracotomy lead configuration when used in conjunction with a device that used biphasic waveforms, 38 consecutive patients were taken to the operating room for implantation of a Cadence tiered therapy defibrillator system. The lead system consisted of a transvenous coil electrode positioned at the right atrial-superior vena caval junction, a bipolar endocardial right ventricular lead, and a large patch placed subcutaneously near the cardiac apex. Of the 38 nonthoracotomy defibrillator implantations attempted, 36 (95%) were completed with adequate defibrillation thresholds. The mean defibrillation threshold in these 36 patients was < or = 563 +/- 10 V (< or = 20 +/- 1 J). There was no perioperative mortality. Complications included coil lead migration (5), sensing lead migration (1), infection (3), pneumothorax (2), arterial embolism (1), and folding of the subcutaneous patch with an increase in defibrillation threshold (1). No patient died during a median follow-up period of 22 weeks. Fourteen patients (39%) had spontaneous sustained ventricular tachyarrhythmias, which were all successfully terminated by the implanted device. Shocks for nonsustained arrhythmias were aborted in eight patients (22%). Spurious discharges for sinus tachycardia or atrial fibrillation occurred in six patients (17%) and were readily diagnosed by examination of the stored electrograms. Thus, implantation of a biphasic tiered therapy defibrillator system using this nonthoracotomy approach is feasible in the majority of patients. The major complication associated with this procedure is lead dislodgment.(ABSTRACT TRUNCATED AT 250 WORDS)Keywords
This publication has 16 references indexed in Scilit:
- Antitachycardia devices: Realities and promisesJournal of the American College of Cardiology, 1991
- Experience with a New Implantable Pacer-, Cardioverter-Defibrillator for the Therapy of Recurrent Sustained Ventricular Tachyarrhythmias: A Step Toward a Universal Ventricular Tachyarrhythmia Control DevicePacing and Clinical Electrophysiology, 1991
- The Initial Clinical Experience with an Implantable Cardioverter Defibrillator/ Antitachycardia PacemakerPacing and Clinical Electrophysiology, 1991
- Improved low energy defibrillation efficacy in man with the use of a biphasic truncated exponential waveformAmerican Heart Journal, 1989
- The automatic implantable cardioverter defibrillator: Efficacy, complications and survival in patients with malignant ventricular arrhythmiasJournal of the American College of Cardiology, 1988
- The automatic implantable cardioverter-defibrillator. Clinical experience, complications, and follow-up in 25 patientsArchives of internal medicine (1960), 1988
- Implantation of a cardioverter/defibrillator without thoracotomy using a triple electrode systemPublished by American Medical Association (AMA) ,1988
- The automatic implantable cardioverter-defibrillator: An overviewJournal of the American College of Cardiology, 1985
- Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator.Circulation, 1985
- Permanent pacemaker infections: Characterization and managementThe American Journal of Cardiology, 1981