Charnley wear model for validation of hip simulators' ball diameter versus polytetrafluoroethylene and polyethylene wear
- 1 January 1997
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 211 (1) , 25-36
- https://doi.org/10.1243/0954411971534656
Abstract
Wear rates of polytetrafluoroethylene (PTFE) and polyethylene cups were compared in 9-channel and 12-channel simulators, using serum lubrication and gravimetric techniques for wear assessment. Cobalt-chromium (CoCr) and alumina ceramic femoral heads in 22-42 mm diameter range were used to validate simulator wear rates against clinical data. This was also the first study of three femoral head sizes evaluated concurrently in a simulator (with three replicate specimens) and also the first report in which any wear experiments were repeated. Fluid absorption artefacts were within ± 1 per cent of wear magnitude for PTFE and ± 8 per cent for polyethylene and were corrected for. Wear rates were linear as a function of test duration. Precision within each set of three cups was within ±6 per cent. The wear rates from experiments repeated over 15 months were reproducible to within ± 24 per cent. However, the magnitudes of the simulator wear rates were not clinically accurate, the PTFE wear rates (2843 mm3/106 cycles; 22 mm diameter) were over three times higher than in vivo, the polyethylene 30 to 50 per cent on the low side (23 mm3/106 cycles; 22 mm diameter). Volumetric wear rate increased with respect to size of femoral head and a linearly increasing relationship of 7-8 per cent/mm was evident with respect to femoral head diameter for both PTFE and polyethylene. These data compared well with the clinical data.Keywords
This publication has 19 references indexed in Scilit:
- Isolation of predominantly submicron‐sized UHMWPE wear particles from periprosthetic tissuesJournal of Biomedical Materials Research, 1995
- Wear of the polyethylene acetabular cup: The effect of head material, head diameter, and cup thickness studied with a hip simulatorActa Orthopaedica, 1995
- Wear of polyethylene acetabular cups against alumina femoral heads: 5 prostheses compared in a hip simulator for 35 million walking cyclesActa Orthopaedica, 1993
- Role of Ceramic ImplantsClinical Orthopaedics and Related Research, 1992
- Fluid-sorption phenomena in sterilized polyethylene acetabular prosthesesBiomaterials, 1985
- Degradation and Wear of Ultra-High-Molecular-Weight PolyethylenePublished by ASTM International ,1985
- Wear of Artificial Joint Materials IVEngineering in Medicine, 1981
- Wear mechanisms for ultrahigh molecular weight polyethylene in the total hip prosthesisJournal of Applied Polymer Science, 1979
- Wear characteristics of UHMW polyethylene: A method for accurately measuring extremely low wear ratesJournal of Biomedical Materials Research, 1978
- Rate of Wear in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975