A Computer-Controlled Double Beam Absorption Spectrophotometer for the Ultraviolet, Visible, and near Infrared Regions
- 1 January 1974
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 28 (1) , 45-51
- https://doi.org/10.1366/000370274774332948
Abstract
A PDP 8/1 computer with 8 K memory has been used to control a double beam absorption spectrophotometer for the range of 50 000 to 10 000 cm−1. Details of the optical and electronic design are given, and the principles and advantages of a computer-controlled instrument are outlined. The software is described, and some spectra demonstrating the performance of the instrument are given.Keywords
This publication has 2 references indexed in Scilit:
- Computerized Treatment of Ultraviolet and Visible Absorption SpectraApplied Spectroscopy, 1974
- Compensation for Light Source Noise in a Sensitive Dual Wavelength SpectrophotometerReview of Scientific Instruments, 1970