False-Positive Immunoassay Results: A Multicenter Survey of Erroneous Immunoassay Results from Assays of 74 Analytes in 10 Donors from 66 Laboratories in Seven Countries
Open Access
- 1 November 2002
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 48 (11) , 2008-2016
- https://doi.org/10.1093/clinchem/48.11.2008
Abstract
Background: Analytical interference in immunoassays can produce serious errors, but it is generally considered rare with modern analytical systems. Method: Blood was collected from 10 donors with illnesses known to be associated with rheumatoid factor. Immunoassays for 74 analytes were performed in 66 clinical laboratories. Each sample was measured in duplicate, and again in duplicate after the addition of a proprietary heterophil blocking reagent, with the laboratory’s routinely used reagents and equipment. Reagents were typically supplied by the manufacturers of the closed analytical systems. Both competitive and sandwich immunoassays were investigated. Results: Overall ∼8.7% of the 3445 results were considered to be “false positive”, many of them seriously so. Twenty-one percent of the erroneously high results (1.8% of all results) were potentially misleading and were corrected by blocking reagent, although 49% of such seriously high results (4.2% of all results) were not. A further 39% of the false-positive results (2.6% of all results) would not necessarily have appeared likely to produce adverse clinical consequences but were substantially lowered by the addition of the blocking reagent. Conclusions: Random errors can occur with all types of immunoassays tested and can be difficult to identify even when repeated in another laboratory. Clinicians need to be aware of these limitations.Keywords
This publication has 19 references indexed in Scilit:
- Automated immunoassay analysersAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2001
- A Mass Spectrometric Method for Quantitation of Intact Insulin in Blood SamplesJournal of Analytical Toxicology, 2001
- Performance of the enhanced Abbott AxSYM Cardiac Troponin I reagent in patients with heterophilic antibodiesClinica Chimica Acta; International Journal of Clinical Chemistry, 2000
- More on Heterophile and Human Anti-Animal AntibodiesClinical Chemistry, 1999
- Interference in ImmunoassayAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1999
- False-Positive Troponin I in the MEIA Due to the Presence of Rheumatoid Factors in Serum: Elimination of This Interference by Using a Polyclonal Antisera Against Rheumatoid FactorsAmerican Journal of Clinical Pathology, 1999
- Antibody multispecificity in immunoassay interferenceClinical Biochemistry, 1992
- Labelled Antibodies and Immunological Assay SystemsNature, 1968
- The estimation of thyroxine in human plasma by an electrophoretic techniqueClinica Chimica Acta; International Journal of Clinical Chemistry, 1960
- IMMUNOASSAY OF ENDOGENOUS PLASMA INSULIN IN MANJournal of Clinical Investigation, 1960