A Quality Control Program for the Radioimmunoassay Laboratory
Open Access
- 1 November 1976
- journal article
- research article
- Published by Oxford University Press (OUP) in American Journal of Clinical Pathology
- Vol. 66 (5) , 854-862
- https://doi.org/10.1093/ajcp/66.5.854
Abstract
The major problems in applying quality control to radioimmunoassay measurements are (a) nonlinearity of the dose-response curve, and (b) nonuniformity of the residual variance. A logit-log transformation of the dose-response variables combined with an iterative weighted regression analysis appears to overcome most of the difficulties. This technic is particularly helpful when applied to substandard runs where extraneous standard points tend, to distort assay results. The authors describe a quality control program that involves recording control values on calendar and histogram formats, monitoring assay variables by charting, and comparing the computer-calculated slope with the graphic plot to reveal “outliers.” This program is useful in guiding technologists to locate possible causes for “out-of-limits” runs.Keywords
This publication has 2 references indexed in Scilit:
- QUANTITATIVE ASPECTS OF THE REACTION BETWEEN INSULIN AND INSULIN-BINDING ANTIBODYJournal of Clinical Investigation, 1959
- INSULIN-I131 METABOLISM IN HUMAN SUBJECTS: DEMONSTRATION OF INSULIN BINDING GLOBULIN IN THE CIRCULATION OF INSULIN TREATED SUBJECTS 1Journal of Clinical Investigation, 1956