Simultaneous Objective Measurements Of Dose And Image Quality In Mammography

Abstract
The performance of a radiological system can be evaluated on the one hand by an objective determination of the quality of the produced image and, on the other hand, by the dose delivered to the patient. In order to measure these two factors in a single exposure a Kodak breast phantom has been modified so as to simulate the breast absorption. The dose distribution is measured with thermoluminescent detectors. By consideration of a theoretical model of the X-ray imaging in mammography, a single quality factor is computed from the contrast, the spatial resolution and the noise measured on the phantom image. We present results obtained in various working conditions, i.e. variable X-ray tube voltages, use of different screen-film combinations, use of a grid.© (1982) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.