Screening for von Willebrand Disease With a New Analyzer Using High Shear Stress: A Study of 60 Cases
Open Access
- 15 February 1998
- journal article
- Published by American Society of Hematology in Blood
- Vol. 91 (4) , 1325-1331
- https://doi.org/10.1182/blood.v91.4.1325
Abstract
We have evaluated the performance of a new analyzer using high shear stress, the PFA-100 (Platelet Function Analyzer, Dade International, Massy, France), for screening of patients with von Willebrand disease (vWD). Whole citrated blood is aspirated through a capillary to the central aperture of a membrane coated with collagen and with a platelet agonist (either epinephrine or adenosine diphosphate [ADP]). The time required to obtain occlusion of the aperture by a platelet plug is defined as the closure time (CT). We studied 60 patients with different types of vWD and 96 normal subjects. Fourteen subjects with hemophilia and 15 patients with a platelet disorder were also analyzed. When omitting results from two patients with type 2N, the 58 other patients with type 1, type 2A, type 2B, type 3, or acquired vWD all exhibited an abnormal occlusion with collagen-ADP (sensitivity, 100%) and 56 of 58 had an abnormal CT with collagen-epinephrine (sensitivity, 96.5%). Only two patients with mild type 1 were not detected with collagen-epinephrine. In comparison, the bleeding time (BT) was normal in 20 patients: 17 with type 1, two with type 2A, and one with acquired vWD (sensitivity, 65.5%). The specificity of the PFA-100 was over 95% with both types of cartridges. Thus, the analyzer is well adapted to routine testing, as it has the advantages of simplicity and ease of execution, and demonstrates a high sensitivity, clearly superior to that of BT, for the screening of patients with vWD.Keywords
This publication has 14 references indexed in Scilit:
- Characterization of an In Vitro Platelet Function Analyzer, PFA-100™Clinical and Applied Thrombosis/hemostasis, 1996
- Preliminary Data from a Field Trial of the PFA-100™ SystemSeminars in Thrombosis and Hemostasis, 1995
- Description of an In Vitro Platelet Function Analyzer-PFA-100™Seminars in Thrombosis and Hemostasis, 1995
- Von Willebrand Disease: A Database of Point Mutations, Insertions, and DeletionsThrombosis and Haemostasis, 1993
- von Willebrand Factor: Structure and FunctionThrombosis and Haemostasis, 1993
- Modulation of platelet function through adhesion receptors. A dual role for glycoprotein IIb-IIIa (integrin alpha IIb beta 3) mediated by fibrinogen and glycoprotein Ib-von Willebrand factor.Journal of Biological Chemistry, 1992
- The Structure and Function of von Willebrand FactorThrombosis and Haemostasis, 1992
- The role of von Willebrand factor and fibrinogen in platelet aggregation under varying shear stress.Journal of Clinical Investigation, 1991
- Fibrinogen-independent platelet adhesion and thrombus formation on subendothelium mediated by glycoprotein IIb-IIIa complex at high shear rate.Journal of Clinical Investigation, 1989
- Relation of von Willebrand Factor to Bleeding TimeNew England Journal of Medicine, 1974