Visual complaints associated with the AcrySof acrylic intraocular lens
- 1 September 2000
- journal article
- case report
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 26 (9) , 1339-1345
- https://doi.org/10.1016/s0886-3350(00)00441-7
Abstract
To describe the visual complaints of a series of patients implanted with the AcrySof(R) (Alcon Surgical) acrylic intraocular lens (IOL) that resolved with IOL exchange. Jules Stein Eye Institute, Los Angeles, California, and John A. Moran Eye Center, Salt Lake City, Utah, USA. This was a retrospective review of patients who had AcrySof IOL exchange from January 1997 to December 1998. Eight patients (9 eyes) with bothersome visual symptoms following AcrySof IOL implantation were identified. Problems included glare, halos around point light sources, and peripheral arcs of light, often worse at night. In each case, the IOL was well-centered in the capsular bag and there was no significant posterior capsule opacification. Six patients (7 eyes) had the MA30BA model with a 5.5 mm optic, and 2 patients (2 eyes) had the MA60BM model with a 6.0 mm optic. No extralenticular reasons for the patients' complaints could be identified. Exchanging the AcrySof IOLs with silicone or poly(methyl methacrylate) IOLs alleviated most symptoms. In 5 of 8 patients, dysphotopsias resolved completely. A small number of patients implanted with AcrySof IOLs have specific complaints of glare, halos, and peripheral arcs of light. Optical considerations that may help explain these symptoms include the high refractive index of the IOL material and the truncated design of the optic. Patients who are highly observant and those with large pupils may be particularly symptomatic. Intraocular lens exchange may be necessary in some cases.Keywords
This publication has 10 references indexed in Scilit:
- Practice styles and preferences of ASCRS members—1999 surveyJournal of Cataract & Refractive Surgery, 2000
- Preventing posterior capsule opacification by creating a discontinuous sharp bend in the capsuleJournal of Cataract & Refractive Surgery, 1999
- Quantitative Comparison of Posterior Capsule Opacification After Polymethylmethacrylate, Silicone, and Soft Acrylic Intraocular Lens ImplantationArchives of Ophthalmology (1950), 1998
- Relationship between intraocular lens biomaterials and posterior capsule opacificationPublished by Wolters Kluwer Health ,1998
- Biocompatibility of poly(methyl methacrylate), silicone, and AcrySof intraocular lenses: Randomized comparison of the cellular reaction on the anterior lens surfaceJournal of Cataract & Refractive Surgery, 1998
- Clinical results of AcrySof intraocular lens implantationJournal of Cataract & Refractive Surgery, 1998
- Cataract surgical problemJournal of Cataract & Refractive Surgery, 1997
- Scanning Electron Microscopic Characteristics of Small-incision Intraocular LensesOphthalmology, 1996
- Visual significance of glistenings seen in the AcrySof intraocular lensJournal of Cataract & Refractive Surgery, 1996
- Two year clinical study of a soft acrylic intraocular lensJournal of Cataract & Refractive Surgery, 1996