An Analysis of Thermal Regulation in Hemodialysis with One and Three Compartment Models
- 1 July 1989
- journal article
- Published by Wolters Kluwer Health in Annual Northeast Bioengineering Conference
- Vol. 35 (3) , 622-624
- https://doi.org/10.1097/00002480-198907000-00148
Abstract
A one compartment model of heat kinetics in hemodialysis (HD) was formulated, solved for the body surface heat transfer coefficient, and used to analyze reported clinical data. The analysis showed a linear relationship between percent change in body surface (y) and fractional dialyzer (x) heat loss, such that y = -25 - 44X, p less than 0.001. Thus, when X = 0, surface heat loss decreases 25% and core temperature (T) rises, and X must equal 45% of heat production to stabilize core T during HD. A three compartment model predicts that skin clothing insulation will have to increase 2.5 times to avoid thermal discomfort when X = 45% of heat production. A theoretic mechanism for decreased symptomatic hypotension resulting from stabilization of core T is proposed.Keywords
This publication has 0 references indexed in Scilit: