Cardiac image reconstruction on a 16-slice CT scanner using a retrospectively ECG-gated multicycle 3D back-projection algorithm
- 16 May 2003
- proceedings article
- Published by SPIE-Intl Soc Optical Eng
- Vol. 5032, 1820-1829
- https://doi.org/10.1117/12.480117
Abstract
Fast 16-slice spiral CT delivers superior cardiac visualization in comparison to older generation 2- to 8-slice scanners due to the combination of high temporal resolution along with isotropic spatial resolution and large coverage. The large beam opening of such scanners necessitates the use of adequate algorithms to avoid cone beam artifacts. We have developed a multi-cycle phase selective 3D back projection reconstruction algorithm that provides excellent temporal and spatial resolution for 16-slice CT cardiac images free of cone beam artifacts.Keywords