First results using automated epifluorescence microscopyto detect Escherichia coli and Staphylococcus epidermidisin WBC‐reduced platelet concentrates
- 24 November 2001
- journal article
- research article
- Published by Wiley in Transfusion
- Vol. 41 (11) , 1351-1355
- https://doi.org/10.1046/j.1537-2995.2001.41111351.x
Abstract
BACKGROUND: Many methods have been tested for the detection of bacterial contamination in platelets. However, only those using molecular biology or cell culturing consistently detect contamination at levels below 105 bacteria per mL. This report describes the initial investigation into an alternative method that offers the possibilities of high sensitivity and rapid response while using available laboratory equipment and supplies. This method relies on a fluorescent nucleic acid stain, which preferentially stains bacteria but not platelets, and automated epifluorescence microscopy for rapid analysis. Measurements in WBC-reduced platelet concentrates (PCs) contaminated with bacteria are reported at concentrations between 103 and 106 bacteria per mL. STUDY DESIGN AND METHODS:Staphylococcus epidermidis or Escherichia coli was inoculated into aliquots of WBC-reduced PCs on Days 2 through 5 of storage. Bacterially inoculated and control PCs were stained, platelets and residual WBCs were lysed, and 200 μL of sample was filtered onto black polycarbonate filters. All preparations were done in triplicate. An automated epifluorescence microscope examined approximately 2 percent of the area of each filter and used image analysis to select the fluorescent particles that should be counted as bacteria. RESULTS: Samples containing 3 to 5 × 103 bacteria per mL produced about three times as many fluorescent particles classified as bacteria as the controls. Lower concentrations of S. epidermidis were detected because of higher fluorescence intensity. Simultaneous preparation of six samples requires about 35 minutes. Analysis of each prepared sample takes 10 minutes, for a total preparation and analysis time of about 95 minutes for 6 samples. CONCLUSION: Low concentrations (3 bacteria/mL) of deliberately inoculated S. epidermidis or E. coli can be measured quickly in WBC-depleted PCs by using a fluorescent nucleic acid stain, differential lysis, and automated microscopy. Continued refinement of the method, studies employing other bacterial strains, and further validations of assay performance are warranted.Keywords
This publication has 25 references indexed in Scilit:
- A retrospective analysis of microbial contaminants in outdated random‐donor platelets from multiple sitesTransfusion, 1997
- Rapid identification of bacterially contaminated platelets using reagent strips: glucose and pH analysis as markers of bacterial metabolismTransfusion, 1997
- Evaluation of swirling, pH, and glucose tests for the detection of bacterial contamination in platelet concentratesTransfusion, 1996
- Bacteria levels in components prepared from deliberately inoculated whole blood held for 8 or 24 hours at 20 to 24 degrees CTransfusion, 1995
- A prospective microbiologic surveillance program to detect and prevent the transfusion of bacterially contaminated plateletsTransfusion, 1993
- The use of a chemiluminescence‐linked universal bacterial ribosomal RNA gene probe and blood gas analysis for the rapid detection of bacterial contamination in white cell‐reduced and nonreduced plateletsTransfusion, 1993
- Strategies for the avoidance of bacterial contamination of blood componentsTransfusion, 1993
- Staphylococcus epidermidis bacteremia from transfusion of contaminated platelets: application of bacterial DNA analysisTransfusion, 1992
- Blood Product-Associated Bacterial SepsisTransfusion Medicine Reviews, 1991
- Evaluation of the buffy-coat smear for rapid detection of bacteremiaPublished by American Medical Association (AMA) ,1981