Characterization of hapten binding to immunoconjugates by electrospray ionization mass spectrometry
- 1 June 1994
- journal article
- Published by American Chemical Society (ACS) in Bioconjugate Chemistry
- Vol. 5 (3) , 194-198
- https://doi.org/10.1021/bc00027a003
Abstract
Electrospray ionization mass spectrometry (ESI-MS) has been used to characterize the covalent binding of different haptens to the enzyme glucose 6-phosphate dehydrogenase. The technique allows one to directly observe the relative amounts of each conjugated species present in a mixture, as well as the average hapten number. These measurements are useful for optimizing reaction conditions to yield a more precisely defined product for use in immunoassays. The results obtained show that ESI-MS with a quadrupole analyzer can be successfully used to analyze mixtures of derivatized proteins in the molecular weight range of 50-60 kDa, where the modifying hapten has a molecular weight as low as 130 Da.Keywords
This publication has 0 references indexed in Scilit: