Geometric misalignment and calibration in cone‐beam tomography
- 15 November 2004
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 31 (12) , 3242-3266
- https://doi.org/10.1118/1.1803792
Abstract
We present a new high-precision method for the geometric calibration in cone-beam computed tomography. It is based on a Fourier analysis of the projection-orbit data, recorded with a flat-panel area detector, of individual point-like objects. For circular scan trajectories the complete set of misalignment parameters which determine the deviation of the detector alignment from the ideal scan geometry are obtained from explicit analytic expressions. To derive these expressions we show how to disentangle the problems of calculating misalignment parameters and point coordinates. The calculation of the coordinates of the point objects inside the scanned volume, in units of the distance from the focal spot to the center of rotation, is then possible analytically likewise. We simulate point-projection data on a misaligned detector with various amounts of randomness added to mimic measurement uncertainties. This data is then employed in our calibration to validate the method by comparing the resulting misalignment parameters and point coordinates to the known true ones. We also present our implementation and results for the geometric calibration of micro-CT systems. The effectiveness of the corresponding misalignment correction in reducing image artifacts is exemplified by reconstructed micro-CT images.Keywords
This publication has 12 references indexed in Scilit:
- Characterization of pinhole SPECT acquisition geometryIEEE Transactions on Medical Imaging, 2003
- Implementation of a cone‐beam reconstruction algorithm for the single‐circle source orbit with embedded misalignment correction using homogeneous coordinatesMedical Physics, 2001
- Alignment of a volumetric tomography systemMedical Physics, 2001
- Enhanced 3-D-reconstruction algorithm for C-arm systems suitable for interventional proceduresIEEE Transactions on Medical Imaging, 2000
- Three-dimensional artifact induced by projection weighting and misalignmentIEEE Transactions on Medical Imaging, 1999
- Virtual alignment of x-ray cone-beam tomography system using two calibration aperture measurementsOptical Engineering, 1999
- Geometrical calibration of x-ray imaging chains for three-dimensional reconstructionComputerized Medical Imaging and Graphics, 1993
- Calculation of the rotational centers in computed tomography sinogramsIEEE Transactions on Nuclear Science, 1990
- Reconstruction for fan beam with an angular‐dependent displaced center‐of‐rotationMedical Physics, 1988
- Estimation of geometrical parameters for fan beam tomographyPhysics in Medicine & Biology, 1987