Evaluation of a Dispensing Instrument (Dynatech MIC-2000) for Preparing Microtiter Antibiotic Plates and Testing Their Potency During Storage
- 1 May 1978
- journal article
- research article
- Published by American Society for Microbiology in Antimicrobial Agents and Chemotherapy
- Vol. 13 (5) , 842-844
- https://doi.org/10.1128/aac.13.5.842
Abstract
The MIC-2000 96-channel dispenser was evaluated for accuracy and repeatability of dispensing 50-μl volumes of water. It was found to perform within the manufacturer's specifications for accuracy of ±5%, but not within the limits for repeatability of ±1%. Overall, instrument performance was found to be satisfactory for the antimicrobial susceptibility testing technique which has been implemented in this laboratory. A method is suggested for simple evaluation of the volumes dispensed. A variety of antibiotics may be stored in Trypticase soy broth in Microtiter plates without loss of potency for periods of at least 2 weeks at −20°C and 4 months at −70°C.This publication has 8 references indexed in Scilit:
- Detailed methodology and implementation of a semiautomated serial dilution microtechnique for antimicrobial susceptibility testing.1970
- An Improved Single-disk Method for Testing the Antibiotic Susceptibility of Rapidly-growing PathogensAmerican Journal of Clinical Pathology, 1970
- Semiautomatic microtechnique for serial dilution-antibiotic sensitivity testing in the clinical laboratory.1968
- Serial Dilution Antibiotic Sensitivity Testing with the Microtitrator SystemAmerican Journal of Clinical Pathology, 1966
- Antibiotic susceptibility testing by a standardized single disk method.1966
- ANTIBIOTIC SUSCEPTIBILITY TESTING BY A STANDARDIZED SINGLE DISK METHOD1966
- AN IMPROVED RAPID METHOD OF TUBE DILUTION ANTIBIOTIC SENSITIVITY TESTING1954
- Microbiological Aspects of Penicillin: I. Methods of Assay.1943