Evaluation and assessment of new disposable strip for determination of plasma potassium concentration.
Open Access
- 1 March 1987
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 40 (3) , 298-301
- https://doi.org/10.1136/jcp.40.3.298
Abstract
Overall laboratory reproducibility for the Ames Seralyzer and plasma potassium test strips was less than 4% within the plasma reference range; and linearity extended from 2.1 to 10.2 mmol/l. Neither bilirubin nor lipaemia interfered in the analysis and selectivity for potassium over sodium was 450:1. Laboratory comparison with automated flame photometry and an indirect ion selective electrode showed a Seralyzer bias of +0.10 (+/- 25D 0.26) and +0.11 (0.27) mmol/l. Against flame photometry, comparison of analyses by six doctors from a coronary care unit and four nurses from a renal dialysis unit showed smaller positive biases (0.06 and 0.02 mmol/l) but substantially larger error ranges (0.43 and 0.55 mmol/l, respectively). Some 49% of the nurses' and 37% of the doctors' results differed from laboratory values by more than 0.2 mmol/l. Potassium test strip results proved acceptably comparable and reproducible when produced by an experienced analyst, but greater variability in clinical hands rendered the system unsuitable for precise monitoring of potassium concentration.This publication has 8 references indexed in Scilit:
- Measurement in Medicine: The Analysis of Method Comparison StudiesJournal of the Royal Statistical Society: Series D (The Statistician), 1983
- Solid-phase colorimetric determination of potassium.Clinical Chemistry, 1982
- Development of dry reagent chemistry for the clinical laboratoryJournal of Analytical Methods in Chemistry, 1981
- Analytical Goals in Clinical Chemistry: Their Relationship to Medical CareAmerican Journal of Clinical Pathology, 1979
- Medical Significance of Laboratory ResultsAmerican Journal of Clinical Pathology, 1968
- A Study of the Accuracy and Precision of Clinical Chemistry Determinations in 170 Canadian LaboratoriesClinical Chemistry, 1963