A new method for measurement of blood pressure, heart rate, and activity in the mouse by radiotelemetry
- 1 May 2000
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 88 (5) , 1537-1544
- https://doi.org/10.1152/jappl.2000.88.5.1537
Abstract
A simple and reliable means for accurate, chronic measurement of pulsatile blood pressure (BP) from conscious, freely moving laboratory mice was developed and validated. The newly developed device consists of a small (1.9 ml, 3.4 g), fully implantable radiotelemetry transmitter. Initial frequency response tests showed an adequate dynamic response; the average −3-dB point found in five transmitters was 145 ± 14 (SD) Hz. BP, heart rate, and locomotor activity were recorded from 16 chronically (30–150 days) implanted mice. Mean arterial and pulse pressure, checked at regular intervals, ranged from 90–140 mmHg and from 30–50 mmHg, respectively, throughout the study. Transmitter BP measurements were validated against a Millar 1.4-Fr. transducer-tipped catheter. The mean error of the transmitters for diastolic pressures was +1.1 ± 6.9 mmHg ( n = 7). The error for systolic pressures was, on average, 2.7 ± 3.9 mmHg larger. This new device accurately monitors BP, heart rate, and locomotor activity in conscious, untethered, freely moving mice living in their home cages for periods of at least 150 days.Keywords
This publication has 7 references indexed in Scilit:
- Long-term measurement of arterial blood pressure in conscious miceAmerican Journal of Physiology-Regulatory, Integrative and Comparative Physiology, 1998
- Noninvasive indexes of cardiac systolic and diastolic function in hyperthyroid and senescent mouseAmerican Journal of Physiology-Heart and Circulatory Physiology, 1996
- A Noninvasive Computerized Tail-Cuff System for Measuring Blood Pressure in MiceHypertension, 1995
- Noninvasive measurement of ascending aortic blood velocity in miceAmerican Journal of Physiology-Heart and Circulatory Physiology, 1995
- Telemetric Monitoring of Cardiovascular Parameters in Conscious Spontaneously Hypertensive RatsJournal of Cardiovascular Pharmacology, 1993
- Use of telemetry to record electrocardiogram and heart rate in freely moving miceJournal of Pharmacological and Toxicological Methods, 1993
- Lack of correlation between direct and indirect measurements of arterial pressure in unanaesthetized ratsCardiovascular Research, 1971