Ultraviolet radiation dosimetry with radiochromic film
- 26 June 2000
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 45 (7) , 1863-1868
- https://doi.org/10.1088/0031-9155/45/7/311
Abstract
Radiochromic film is tested for its broad-band response to ultraviolet (UV) B (290-320 nm) and A (320 nm-400 nm), visible and infrared radiation produced by a solar simulator and examined for dosimetry in ultraviolet radiation. Results show that MD-55-2 radiochromic film in solar and fluorescent light sources responds almost exclusively to broad-band UVA radiation with negligible colouration from UVB, visible and low level infrared radiation. A second order polynomial function approximates the change in optical density at 660 nm wavelength for film colouration with exposure to UVA from white light fluorescent and solar UV with exposures measured with a dedicated UVA dosimeter. Using a double exposure technique as used in radiation dosimetry where the film is firstly irradiated to a known UV dose, radiochromic film can be used as a quantitative measure of UVA exposure.Keywords
This publication has 14 references indexed in Scilit:
- Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55Medical Physics, 1998
- Measurement of off-axis and peripheral skin dose using radiochromic filmPhysics in Medicine & Biology, 1998
- Quantification of personal solar UV exposure of outdoor workers, indoor workers and adolescents at two locations in Southeast QueenslandPhotodermatology, Photoimmunology & Photomedicine, 1998
- An Estimate of the Total UV-B Exposure for Outdoor Workers During a South–East Queensland SummerHealth Physics, 1997
- Dosimetric characteristics of an improved radiochromic filmMedical Physics, 1996
- Effects of ultraviolet irradiation on human skin‐derived epidermal cells in vitroJournal of Cellular Physiology, 1993
- Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetryMedical Physics, 1991
- Sensitometry of the response of a new radiochromic film dosimeter to gamma radiation and electron beamsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1991
- THE SUNBURNING ULTRAVIOLET METER: DESIGN AND PERFORMANCEPhotochemistry and Photobiology, 1976