Validation of OPS imaging for microvascular measurements during isovolumic hemodilution and low hematocrits
- 1 April 2002
- journal article
- Published by American Physiological Society in American Journal of Physiology-Heart and Circulatory Physiology
- Vol. 282 (4) , H1502-H1509
- https://doi.org/10.1152/ajpheart.00475.2001
Abstract
Orthogonal polarization spectral (OPS) imaging is a new technique that can be used to visualize the microcirculation with reflected light. It uses hemoglobin absorption to visualize the red blood cells (RBCs). Thus the method could fail at low hematocrit (Hct). The aim of this study was to validate OPS imaging for quantitative measurements of diameter and functional capillary density (FCD) under conditions of hemodilution of varying degrees to achieve a wide range of Hcts. The validation was performed in the dorsal skinfold chamber of nine awake Syrian golden hamsters. Measurements of vessel diameter and FCD were performed off-line using Cap-Image on video sequences captured using OPS imaging and standard intravital fluorescence microscopy at baseline, 85, 70, 55, and 40% of the initial Hct. For hemodilution, isovolumic exchange of blood for 6% Dextran 60 was performed. Bland-Altman plots for the vessel diameter and FCD show good agreement between the two methods for both parameters at all studied Hcts. As expected, there was a systematic bias of ∼4 μm in the diameter measurements since the RBC column was measured and not the intravascular diameter. In conclusion, OPS imaging can be used to measure diameter and FCD at a wide range of Hcts.Keywords
This publication has 13 references indexed in Scilit:
- The CytoscanTM Model E-II, a New Reflectance Microscope for Intravital Microscopy: Comparison with the Standard Fluorescence MethodJournal of Vascular Research, 2000
- Orthogonal polarization spectral imaging: A new method for study of the microcirculationNature Medicine, 1999
- Cap Image - ein neuartiges computerunterstütztes Videobildanalysesystem für die dynamische Kapillarmikroskopie - Cap Image - A Newly-developed Computer-aided Videoframe Analysis System for Dynamic CapillaroscopyBiomedizinische Technik/Biomedical Engineering, 1997
- Comparing methods of measurement: why plotting difference against standard method is misleadingThe Lancet, 1995
- Functional Capillary Density: An Indicator of Tissue Perfusion?International Journal of Microcirculation, 1995
- Measurement Uncertainties Associated with the Use of Bright-Field and Fluorescence Microscopy in the MicrocirculationMicrovascular Research, 1995
- Tissue Oxygenation after Prolonged Ischemia in Skeletal Muscle: Therapeutic Effect of Prophylactic Isovolemic HemodilutionPublished by Springer Nature ,1989
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Technical report—a new chamber technique for microvascular studies in unanesthetized hamstersZeitschrift für Die Gesamte Experimentelle Medizin, 1980
- Estimating Allowable HemodilutionAnesthesiology, 1974