Fatigue testing and performance of acrylic bone‐cement materials: State‐of‐the‐art review
- 5 June 2003
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research Part B: Applied Biomaterials
- Vol. 66B (1) , 457-486
- https://doi.org/10.1002/jbm.b.10018
Abstract
Over the past three decades or so, a very large volume of literature has been generated on the impact of an assortment of variables on the fatigue lifetimes of a large number of acrylic bone‐cement formulations. In the present article, this literature is examined critically to reveal areas of agreement, areas of disagreement, as well as a welter of underexplored and unexplored topics. For example, there is unanimity of support for the notion that an increase in the molecular weight of the powder constituents or the fully cured cement leads to an increase in the cement's fatigue life, whereas there is disagreement as to whether vacuum mixing the cement constituents leads to an increase in the fatigue life of the fully cured cement (relative to the hand‐mixed counterpart). Among the underexplored topics is systematic study of the effect of test frequency on the fatigue results, whereas determination of the optimal concentration of the antibiotic in an antibiotic‐loaded cement is an example of the unexplored topics. It is pointed out that resolving the controversies, addressing the underexplored topics, and filling the lacunae will allow comprehensive evaluations of acrylic bone‐cement materials to be made. This enhanced body of knowledge will prove invaluable in the continued use of acrylic bone cement as the anchoring agent in cemented arthroplasties. © 2003 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 66B: 457–486, 2003Keywords
This publication has 80 references indexed in Scilit:
- The type of cement and failure of total hip replacementsThe Journal of Bone and Joint Surgery. British volume, 2002
- Estimation of Weibull parameters for the flexural strength of PMMA‐based bone cementsPolymer Engineering & Science, 2002
- Mechanical properties of acrylic bone cement containing PMMA-SiO2 hybrid sol-gel materialJournal of Biomedical Materials Research, 1997
- Low-vs high-viscosity cement in hip arthroplasty: No radiographic difference in 226 arthrosis cases followed for 5 yearsActa Orthopaedica, 1993
- Fatigue failure in the cement mantle of an artificial hip jointClinical Materials, 1993
- Physical characterization of acrylic bone cement cured with new accelerator systemsClinical Materials, 1991
- Fatigue testing of acrylic bone cements: Statistical concepts and proposed test methodologyJournal of Biomedical Materials Research, 1988
- Tensile Fatigue Failure of Acrylic Bone CementJournal of Biomechanical Engineering, 1983
- Fracture and fatigue characteristics of bone cementsJournal of Biomedical Materials Research, 1979
- Fracture characteristics of acrylic bone cements. I. Fracture toughnessJournal of Biomedical Materials Research, 1976