Assessment of motion gating strategies for mouse magnetic resonance at high magnetic fields
Open Access
- 21 January 2004
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 19 (2) , 229-237
- https://doi.org/10.1002/jmri.10454
Abstract
Purpose To assess the performance of motion gating strategies for mouse cardiac magnetic resonance (MR) at high magnetic fields by quantifying the levels of motion artifact observed in images and spectra in vivo. Materials and Methods MR imaging (MRI) of the heart, diaphragm, and liver; MR angiography of the aortic arch; and slice‐selective 1H‐spectroscopy of the heart were performed on anesthetized C57Bl/6 mice at 11.75 T. Gating signals were derived using a custom‐built physiological motion gating device, and the gating strategies considered were no gating, cardiac gating, conventional gating (i.e., blanking during respiration), automatic gating, and user‐defined gating. Both automatic and user‐defined modes used cardiac and respiratory gating with steady‐state maintenance during respiration. Gating performance was assessed by quantifying the levels of motion artifact observed in images and the degree of amplitude and phase stability in spectra. Results User‐defined gating with steady‐state maintenance during respiration gave the best performance for mouse cardiac imaging, angiography, and spectroscopy, with a threefold increase in signal intensity and a sixfold reduction in noise intensity compared to cardiac gating only. Conclusion Physiological gating with steady‐state maintenance during respiration is essential for mouse cardiac MR at high magnetic fields. J. Magn. Reson. Imaging 2004;19:229–237.Keywords
This publication has 27 references indexed in Scilit:
- Navigator-echo-based real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography.Radiology, 1996
- Respiratory Blur in 3D Coronary MR ImagingMagnetic Resonance in Medicine, 1995
- High resolution high field rodent cardiac imaging with flow enhancement suppressionMagnetic Resonance Imaging, 1994
- Controlled ventilation during NMR spectroscopic studies: Hemodynamic and biochemical consequencesMagnetic Resonance Imaging, 1991
- Adaptive technique for high-definition MR imaging of moving structures.Radiology, 1989
- Motional phase artifacts in fourier transform MRIMagnetic Resonance in Medicine, 1989
- The magnetic field dependence of the breathing artifactMagnetic Resonance Imaging, 1986
- Respiratory effects in two-dimensional Fourier transform MR imaging.Radiology, 1986
- Physiological monitoring during nmr measurements of animalsJournal of Magnetic Resonance (1969), 1983
- The signal-to-noise ratio of the nuclear magnetic resonance experimentJournal of Magnetic Resonance (1969), 1976