Clinical and Hip Simulator Comparisons of Ceramic-on-Polyethylene and Metal-on-Polyethylene Wear
- 1 October 2000
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 379 (379) , 34-40
- https://doi.org/10.1097/00003086-200010000-00006
Abstract
The benefit of reduced polyethylene wear with ceramic in hip replacements does not seem to have been universally appreciated. In this current study, wear predictions from laboratory and clinical studies were compared for ceramic-onpolyethylene and cobalt chrome-on-polyethylene combinations. Many laboratory studies included water-based lubrication and linear-tracking mechanisms. Now it is appreciated that these were inappropriate methods, because of a propensity for very low or virtually no polyethylene wear against ceramics in water. Thus, water-based studies predicting a 20-to 80-fold advantage for ceramic-on-polyethylene compared with metal-on-polyethylene clearly were in error. However, serum-based simulator studies with high protein-concentrations generally have shown greater wear with alumina-on-polyethylene than with metal-on-polyethylene. Controversy still remains over the use of such nonphysiologic protein levels. The simulator studies were just beginning to explore the role of serum protein concentrations and the influence on the various wear models. Polyethylene wear with zirconia systems was particularly affected by serum protein concentrations. In one simulator study, use of proteins in the physiologic range resulted in the alumina-on-polyethylene wear rate decreasing to approximately 50% of that of metal-on-polyethylene. In the literature, many hip design and polyethylene variations were reported which confounded the wear analysis. Overall, the clinical data supported the superior performance of ceramic-on-polyethylene systems by a factor of 1.5-to fourfold. However, the amount of supporting data was not large. This summary of laboratory and clinical data indicated that ceramic-on-polyethylene hip replacement systems offered on average a 50% wear reduction from metal-on-polyethylene systems.Keywords
This publication has 23 references indexed in Scilit:
- Wear simulation of UHMWPE for total hip replacement with a multidirectional motion pin-on-disk device: Effects of counterface material, contact area, and lubricantJournal of Biomedical Materials Research, 1999
- The Significance of Nonlinear Motion in the Wear Screening of Orthopaedic Implant MaterialsJournal of Testing and Evaluation, 1997
- Comparison of the size and morphology of UHMWPE wear debris produced by a hip joint simulator under serum and water lubricated conditionsBiomaterials, 1996
- The effect of transfer film and surface roughness on the wear of lubricated ultra-high molecular weight polyethyleneClinical Materials, 1993
- Prognosis of total hip replacement in Sweden: Follow-up of 92,675 operations performed 1978–1990Acta Orthopaedica, 1993
- Wear of the polyethylene acetabular cup: Metallic and ceramic heads compared in a hip simulatorActa Orthopaedica, 1993
- Ten-year socket wear in 66 hip arthroplasties: Ceramic versus metal headsActa Orthopaedica, 1990
- Friction and wear properties of polymer, metal, and ceramic prosthetic joint materials evaluated on a multichannel screening deviceJournal of Biomedical Materials Research, 1981
- Wear characteristics of UHMW polyethylene: A method for accurately measuring extremely low wear ratesJournal of Biomedical Materials Research, 1978
- New prospects for a prolonged functional life‐span of artificial hip joints by using the material combination polyethylene/aluminium oxide ceramic/metalJournal of Biomedical Materials Research, 1977