A Screened Emitter Photoelectron Analyzer
- 1 December 1970
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 41 (12) , 1740-1742
- https://doi.org/10.1063/1.1684400
Abstract
A retarding field energy analyzer for photoelectron spectroscopy is improved by adding a screened field‐free region around the emitter. For preliminary measurements on nickel, the analyzer error was estimated to be <2.8% of the kinetic energy. The electronic circuit configuration which performed best is presented along with a description of spurious effects introduced by circuit variations.Keywords