Sudden Death Mortality in Implantable Cardioverter Defibrillator Patients
- 1 February 1991
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 14 (2) , 250-254
- https://doi.org/10.1111/j.1540-8159.1991.tb05102.x
Abstract
Implanfable Cardioverter defibrillator (ICD) prevention of sudden cardiac death (SCD) is not absolute and our experience was reviewed to determine the frequency and nature of SCO in this population. The incidence and cause of mortality in 56 consecutive patients, who underwent ICD implantation beginning May 1982 with follow‐up through May 19, 1990 were analyzed. Twenty‐one patients died, 33% of the mortality was due to SCD, and 52% of deaths may be considered arrhythmic. The cumulative 1, 3, and 5 year SCD survivals were 93%, 89%, and 75%. All seven patients dying of SCD presented initially with SCD, all received previous shocks prior to SCD, and two of the seven patients had devices that were probably inactive at the time of death. We conclude that ICDs reduce but by no means eliminate arrhythmic death, particularly in those at highest risk for SCD. Arrhythmic death remained the most common cause of death in this population.Keywords
This publication has 6 references indexed in Scilit:
- Actuarial Incidence and Pattern of Occurrence of Shocks Following Implantation of the Automatic Implantable Cardioverter DefibrillatorPacing and Clinical Electrophysiology, 1989
- Drug Therapy for Survivors of Sudden Cardiac DeathPacing and Clinical Electrophysiology, 1988
- Implantable Defibrillation: Eight Years Clinical ExperiencePacing and Clinical Electrophysiology, 1988
- Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator.Circulation, 1985
- The Automatic Implantable Cardioverter‐DefibrillatorPacing and Clinical Electrophysiology, 1984
- Determinants of Survival in Patients with Ventricular TachyarrhythmiasNew England Journal of Medicine, 1983