Implantable loop recorder undersensing mimicking complete heart block
- 1 April 2002
- journal article
- case report
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 4 (2) , 211-213
- https://doi.org/10.1053/eupc.2002.0229
Abstract
The newer implantable loop recorders recently introduced (Reveal Plus, Medtronic Inc, Minneapolis, MN, U.S.A.), equipped with auto activation capabilities, are expected to expand our diagnostic abilities in the investigation of syncopal episodes. The case presented here is that of a man with pre-syncopal and syncopal episodes, without organic heart disease, in whom the device was auto-activated due to undersensing and the recorded strip imitated complete heart block. Increased attention should be given by the interpreter of the device's recordings to avoid false positive diagnoses that might lead to serious therapeutic actions.Keywords
This publication has 5 references indexed in Scilit:
- How Can We Identify the Best Implantation Site for an ECG Event Recorder?Pacing and Clinical Electrophysiology, 2000
- Insertable loop recorder in unmasking the cause of syncope.Pacing and Clinical Electrophysiology, 2000
- Use of an Extended Monitoring Strategy in Patients With Problematic SyncopeCirculation, 1999
- Evaluation and Management of the Patient With SyncopeJAMA, 1992
- Prospective evaluation of syncopeAnnals of Emergency Medicine, 1984